CIT-8920 Inductive conductivity manual V1.4
+
  • CIT-8920 Inductive conductivity manual V1.4

CIT-8920 Inductive conductivity manual V1.4

CIT-8920 Inductive Conductivity /Concentration Transmitter/Controller is a high-precision, cost-effective instrument that integrates conductivity/concentration/TDS/temperature measurement.

Keywords:

Product details

Application
• CIP Pipeline Online Cleaning System for the Beverage, Brewing and Dairy Industry;
• Online monitoring of conductivity of highly salty or acid-base corrosive concentrated electrolyte aqueous solutions;
• Metal refining, metal surface treatment and mining industry
• Monitoring of Water Quality in Petrochemical, Electric, Pharmaceutical, Chemical, Water Treatment, Semiconductor Manufacturing Industry;
General
CIT-8920 Inductive Conductivity /Concentration Transmitter/Controller is a high-precision, cost-effective instrument that integrates conductivity/concentration/TDS/temperature measurement.
It is an electromagnetic induction conductivity sensor that is used to measure the conductivity of conductive ions in strong electrolytes. Or a remote transmitter of percent concentration, suitable for continuous measurement of conductivity or percent concentration in a strong ion concentration environment;
The inductive sensor has the anti-pollution ability to contain oil pollution and process coating medium. The sensor has no plate directly contacting the test solution, and there is no polarization
phenomenon, and is not affected by the coating or film form;
The material of the sensor has the advantages of anti-pollution, high temperature resistance and corrosion resistance;
The transmitter has a local display function, using a 128x64 resolution graphic dot matrix liquid crystal display as a human-computer interaction interface, a pure white background backlight,
and a conductivity/temperature/transmission parameter on the same screen.
Suitable for various installations such as pipelines and submersibles required for industrial production. The sensor also has a variety of installation methods and accessories, suitable for different industrial sites.
Technical Features
 
Model
CIT-8920
Product
Inductive Conductivity/Concentration Transfer Controller
Measurement
Conductivity
500μS/cm ~2000mS/cm
Concentration
1.NaOH,(0-15)% or(25-50)%;
2.HNO3(note the Corrosion resistance of the sensor)
(0-25)% or(36-82)%;
3.User-defined concentration curves.
TDS
250.0ppm ~ 1000ppt
Temperature
(0 ~ 120)℃
Resolution
Conductivity
0.01μS/cm
Concentration
0.01%
TDS
0.01ppm
Temperature
0.1℃
Accuracy
Conductivity
(500~1999)μS/cm,±1.5%(FS)
(2~19.9) mS/cm,±1.0% (FS)
(20~199)mS/cm,±1.0% (FS)
(200~2000)mS/cm,±1.0% (FS)
TDS
1.5 level
Temperature
±0.5℃
Temperature
Element
Pt1000
Range
(0~120)℃ linear compensation
Current output
Isolated (4 ~ 20) mA output, maximum loop resistance 500Ω
Accuracy: ± 0.1mA
Control output
Semiconductor Photoelectric Relay,
Load Current 50mA (Max), AC/DC 30V (Max)
Communication
RS485, (please check the protocol )
Power supply
DC 24V±4V
Max consumption
≤2.5W
Protection level
IP65
Working environment
Temperature: (0~50) ℃ Relative Humidity:≤95%RH(non-condensing )
Storage
Temperature:(-20~60) ℃ Relative Humidity:≤85%RH(non-condensing)
Installation
Pipeline mounted /Flow device /Immersion