Digital sensor CODuv-6000 COD Operation manual  V1.0
+
  • Digital sensor CODuv-6000 COD Operation manual  V1.0

Digital sensor CODuv-6000 COD Operation manual V1.0

Directly used as the data acquisition terminal of DCS and PLC system in the industrial field;

Category:

Keywords:

Product details

Applications

• Directly used as the data acquisition terminal of DCS and PLC system in the industrial field;

• Direct access to the cloud communication and big data platform through the DTU of the GPRS system;

• Directly use the WiFi module as a nearby wireless data collection terminal;

• Can use ZigBee/Lora module to remote wireless collection terminal;

• It can realize optical remote data collection through optical transceiver and far-infrared terminal;

• Supporting traditional 4-20mA industrial system (cable end RS485/4-20mA) module conversion;

• With a variety of 485/WiFi/USB tool converter, you can achieve calibration, address search, baud rate search and configuration by APP.

 

 

Diagram 4.1 Different display mode of data acquisition terminal

 

 

4.2 Connection diagram

General Introduction

The CODUV-6000 data collection terminal is an online detection instrument that uses optical methods to detect COD content in water quality. The detection process does not require chemical reagents, obtains data quickly, and has no secondary pollution, so it is more environmentally friendly and has good application prospects. The instrument is embedded with an intelligent core of measurement and calculation, and directly outputs the COD value through RS485 digital signal. It can also convert RS485 digital signal into RS232, 4-20mA, 4G, GPRS, Zigbee, LORA and other data formats through the data conversion module.

Technical index

Product Name
CODUV-6000 Data Acquisition Terminal
Model
CODUV-6001
CODUV-6002
CODUV-6003
CODUV-6004
Measurement
range
0-1000mg/L
0-400mg/L
0-75mg/L
0-3000mg/L
Accuracy
(based on KHP)
Less than ±
5% of reading
or 2mg/L
Less than ±
5% of reading
or 1mg/L
Less than ±
5% of reading
or 0.5mg/L
Less than ±
5% of reading
or 5mg/L
Repeatability
±5% of reading
Working
environment
Temperature:(0~45)℃ ;Humidity:≤95%RH(none considation)
Pressure
0.4Mpa
Storage
environment
Temperature:(-10~60)℃ ;Humidity:≤95%RH(none considation)
Cable length
Standard 10m(out of factory);support(1~300)m(book)
Communication
RS485 communication Modbus-RTUstandard protocol
Power supply
DC(9-28)V
Power
<5W
Weight
1.5Kg
Protection
level
IP68

 

Workshop

Online inquiry


Product name:

Submit