CLA-7000 series  Free  chlorine (DPD) Online
+
  • CLA-7000 series  Free  chlorine (DPD) Online

CLA-7000 series Free chlorine (DPD) Online

Free Chlorine (DPD) online automatic analyzer is mainly used for real-time monitoring of water plants, secondary water supply of pipe network, medical wastewater, kidney dialysis medical water, food and beverage water, swimming pools, circulating cooling water, enterprise discharge wastewater and other occasions. According to the complex and changeable working conditions at the site, it is recommended that the water sample containing suspended solids be equipped with a sample supply filter system to meet the process requirements of different industries and different clients.

Keywords:

Product details

Instrument application
Free Chlorine (DPD) online automatic analyzer is mainly used for real-time monitoring of water plants, secondary water supply of pipe network, medical wastewater, kidney dialysis medical water, food and beverage water, swimming pools, circulating cooling water, enterprise discharge wastewater and other occasions. According to the complex and changeable working conditions at the site, it is recommended that the water sample containing suspended solids be equipped with a sample supply filter system to meet the process requirements of different industries and different clients.
Overview
Free Chlorine (DPD) online automatic analyzer is an online analytical instrument that uses photoelectric colorimetry for the detection of free chlorine disinfectants, which conforms to the principle of national standard free chlorine measurement. Free chlorine (DPD) online automatic analyzer has the characteristics of simple structure, easy maintenance, long maintenance intervals, accurate test data, and good stability. It can be used in medical wastewater, swimming pool water quality management, pipe network water and other occasions. At the same time, to provide you with better application solutions,, we have developed a water sample supply system that makes the operation of the instrument more stable through a large number of operating practices, so that the online analytical instrument can operate accurately and stably.
Instrument measurement principle:
Under weakly acidic conditions, free chlorine in water reacts with DPD reagent to produce red compounds. At a certain wavelength, the absorbance is measured by spectrophotometry and converted into free chlorine value based on the absorbance.
Technical specifications
 
Measurement range
N,N-Diethyl-1,4-phenylenediamine (DPD) spectrophotometry
Model
CLA-7112
CLA-7212
CLA-7113
CLA-7213
Inlet channel
Single channel
Double channel
Single channel
Double channel
Measurement range
Free chlorine:(0.0~
2.0)mg/L ,Calculated as Cl2;
Free chlorine:(0.5~
10.0)mg/L ,Calculated as Cl2;
pH:(0-14);Temperature:(0-100)℃
Accuracy
Free chlorine:±10% or ±0.05mg/L(take the large value),Calculated as Cl2;
Free chlorine:±10% or±0.25mg/L(take the large value),Calculated as Cl2;
pH:±0.1pH;Temperature:±0.5℃
Measurement Period
≤2.5min
Sampling interval
The interval (1~999) min can be set arbitrarily
Maintenance cycle
Recommended once a month (see maintenance chapter)
Environmental requirements
A ventilated and dry room without strong vibration;Recommended room temperature:(15~28)℃;Relative humidity:≤85%(No condensation)
Water sample flow
(200-400) mL/min
Inlet pressure
(0.1-0.3) bar
Inlet water temperature range
(0-40)℃
Power supply
AC (100-240)V; 50/60Hz
Power
120W
Power connection
The 3-core power cord with plug is connected to the mains socket with ground wire
Data output
RS232/RS485/(4~20)mA
Size
H*W*D:(800*400*200)mm

Workshop

Online inquiry


Product name:

Submit