ROS-7000 Reactive Oxygen Species Online automatic
+
  • ROS-7000 Reactive Oxygen Species Online automatic

ROS-7000 Reactive Oxygen Species Online automatic

The Reactive Oxygen Species Online automatic analyzer is mainly used for real-time monitoring of water plants disinfected with potassium monopersulfate disinfectant. According to the complex and changeable working conditions on site, it is recommended that the water sample containing suspended solids be equipped with a sample supply filter system to meet the process requirements of different industries and different customers.

Category:

Keywords:

Product details

Instrument application
The Reactive Oxygen Species Online automatic analyzer is mainly used for real-time monitoring of water plants disinfected with potassium monopersulfate disinfectant. According to the complex and changeable working conditions on site, it is recommended that the water sample containing suspended solids be equipped with a sample supply filter system to meet the process requirements of different industries and different customers.
Overview
ROS-7000 Reactive Oxygen Species Online automatic analyzer is an online analysis instrument that uses photoelectric colorimetry to detect Potassium monopersulphate disinfectants. ROS-7000 Reactive Oxygen Species Online automatic analyzer has the characteristics of simple structure, easy maintenance, long maintenance intervals, accurate test data, and good stability. It can be applied to industries such as waterworks
and end of pipe network. At the same time, through a large number of operating practices, a water sample supply system that makes the operation of the instrument more stable has been introduced to provide you with
better application solutions, so that the online analytical instrument can operate accurately and stably.
Instrument measurement principle:
The water sample reacts with the reagent to produce a red compound. At a certain wavelength, the absorbance is measured and converted into active oxygen based on the absorbance.
Main performance index
 
Measuring Method
Spectrophotometry
Model
ROS-7101
ROS-7201
Inlet water channel
Single channel
Dual channel
Measurement range
(0.0~1.0)mg/L
(optional)pH:(0-14);Temp.:(0-100)℃
Accuracy
±5% or ±0.05mg/L(take the large value )
pH:±0.1pH;Temp.:±0.5℃
Measurement cycle
≤3min
Sampling interval
The interval (1~999) min can be set to any value
Maintenance cycle
Recommended once a month (see maintenance chapter)
Environmental
requirements
Ventilated and dry room without strong vibration;
Suggested room temperature: (15 ~ 28)℃; relative humidity: ≤85% (no
condensation).
Sample water flow
(200-400) mL/min
inlet water pressure
(0.1-0.3) bar
Inlet water temperature range
(0-40)℃
Power supply
AC (100-240)V; 50/60Hz
Consumption
120W
Power connection
3-core power cord with plug is connected to the mains socket with ground wire
Data output
RS232/RS485/(4~20)mA
Dimension size
H*W*D:(800*400*200)mm

 