NH3-N-1400 Ammonina instruction
Category:
Keywords:
Product details
Application
Test sample liquid can be measured without any pretreatment,the sample water can riser directly into the system to measure the ammonia concentration. The maximum measurement range of this equipment is (0 ~ 300.0) mg / L NH3-N. The method is mainly used for on-line automatic monitoring of ammonia nitrogen concentration such as point source and surface water.
General
Please carefully read this manual before boot operation and operate in strict accordance with the instructions. Especially pay attention to all instructions on risk and cautious problem. Please don't repair or do disassemble any components without authorization. Otherwise it may cause serious injury to operators and serious damage to the instrument. For disassembling instruments by yourself, the company will no longer provide free services.
Please do not install the analyzer immediately ,Hold on one day from the transport arrival date .Please use the front hand tighten the black connector to prevent loose joints .
Technical Specification
Range of measurement: (0~ 300.0) mg/L NH3-N
Accuracy: Indication error: S2.0mg/L : +0.2mg/L ;>2.0mg/L: +10%.
Repeatability: less than +3%;
Measurement period: the minimum measurement period is 25 minutes
Sampling period: time interval ((1 ~ 999) min arbitrary and adjustable ) and hourly measuring model.
Calibrating period: Automatically manual calibrated at any time or set the cycle of automatic calibration.
Output: (4~20) mA current output
Maintenance period: Usually monthly, about 30 mins every time.
Environmental requirement:adjustable indoor temperature, proposal temperature(+5~ 40)C humidity≤90% (no condensation).
Power Supply: AC (220+10%) V,(50士10%) Hz, 10A
Dimension: Length 550xwidth 400xheight 1500mm
Other: data is not lost for abnormal alarm and power failure;
Touch screen display and input instructions;
If abnormal reset and power supply after power failure, the instrument will atomatically discharge
the residual reactants and automatically return to work status.